Rohde & Schwarz VOR/ILS Analyzer EVS200 Instrukcja Użytkownika Strona 62

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 68
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 61
VOR/ILS Analyzer EVS200
0796.1800.02 E-8 53
Chapter 4: Interfaces
Interfaces
Antenna input
labeling figure description
RF INPUT 50 input level: 0 dBm / 50
max. +15 dBm
(Data safety up to 10 dBm is ensured)
frequency range: 74 to 341 MHz
VSWR: <1.5
AF output
labeling figure description
AF OUT 50 output level: 200 mV
rms
/50
frequency range: 0.3 to 3.4 kHz
Headphone connection
labeling figure description
PHONES
Only for connecting headphones
with soundproofing.
e.g. R&S order no. 0708.9010.00
XY-Tracer connection
labeling figure description
DSP-OUT ThescalingoftheXYvaluescanbesetin
the setup / Ils mode (Y / t setup) for the
localizer and glideslope mode.